Vertical Autoclave/Steam sterilizers IVA-1000 Series
Vertical Autoclave/Steam sterilizers IVA-1000 Series are floor standing self-contained sterilizer units that require no plumbing and operate from a standard single-phase power supply. Equipped with an effective air-cooling system for liquid sterilisation. This speeds cooling time by up to 40% compared to standard models. The Iris Vertical Autoclaves are available from 36 litres, 54 litres, 80 litres, and 100 litres chamber capacity. All are built with castor wheels as standard making them easy to move. The IVA-1000 is widely used for sterilisation applications in laboratory, hospitals, and clinics as well as for industr
Description
Specification
Technical
Vertical Autoclave/Steam sterilizers IVA-1000 Series are floor standing self-contained sterilizer units that require no plumbing and operate from a standard single-phase power supply. Equipped with an effective air-cooling system for liquid sterilisation. This speeds cooling time by up to 40% compared to standard models. The Iris Vertical Autoclaves are available from 36 litres, 54 litres, 80 litres, and 100 litres chamber capacity. All are built with castor wheels as standard making them easy to move. The IVA-1000 is widely used for sterilisation applications in laboratory, hospitals, and clinics as well as for industrial sterilisation applications in biotechnology and pharmaceutical industries. Comes with sterilization cycle including solid cycle, Liquid cycle, Liquid with warming cycle, Agar cycle, Solid with drying (for D model), Customized cycle.
Adopted microcomputer control system
Comes with Multilayered pressure protection system with automatic interlocking device
Real-time pressure changes monitoring
Comes with double over pressure protection with safety valve and a pressure protective device
Unique anti-scald design
Provide 5 sterilization cycles for sterilization, drying, warming of solid and liquid articles and for melting and warming of agar separately
Temperature and pressure: 105 to 135℃,designed pressure:0.3Mpa
Automatic troubleshooting system
Provide safety with self-induction pressure interlocking device, lid closing checking, over temperature protection, temperature monitor, dry scorch protection system, over pressure protection, safety valve, over current and short circuit protection, leakage protection device, anti-scald safety protection, automatic troubleshooting
Optional Accessories
Stainless steel baskets
Printer
printing set (pressure sensor + printer)
stainless steel pails
Model
IVA-1200
IVA-1200D
IVA-1400
IVA-1400D
IVA-1600
IVA-1600D
IVA-1800
IVA-1800D
Difference
Without Drying
With Drying
Without Drying
With Drying
Without Drying
With Drying
Without Drying
With Drying
Capacity
36 L
54L
80 L
100 L
Materials
SUS 304
Sterilization Cycle
Solid cycle, Liquid cycle, Liquid with warming cycle, Agar cycle, Solid with drying (for D model), Customized cycle
Temperature Accuracy
0.1℃
Melting Temperature
60-100℃
Melting Time
1 to 6000 minutes
Operating Pressure
0.25Mpa
Maximum Pressure
0.3Mpa
Steel Basket
2x(300x200 mm)
3x(300x200 mm)
2x(360x280 mm)
3x(360x220) mm
Sterilization Temperature
105-135℃
Sterilization Time Range
1 to 6000 minutes
Drying Temperature
80 to 140℃
Display
LCD
Drying Time Range
1 to 300 minutes
Working Temperature
5 to 45℃
Humidity
10 to 85%
Certification
-
-
-
-
-
-
-
-
Storage
60 cycles, record 1000 process
Power Consumption
2300 W
2600 W
4600 W
4600 W
Dimension
460x542x1070 mm
460x542x1070 mm
620x680x1100 mm
620x680x1200 mm
Net Weight
Electrical Requirement
220V,16A,50/60HZ
220V,32A,50/60HZ
Catalogue No.
580400502
580401002
580401502
580402002
580402502
580403002
580403502
580404002
Model
IVA-1200
IVA-1200D
IVA-1400
IVA-1400D
IVA-1600
IVA-1600D
IVA-1800
IVA-1800D
Difference
Without Drying
With Drying
Without Drying
With Drying
Without Drying
With Drying
Without Drying
With Drying
Capacity
36 L
54L
80 L
100 L
Materials
SUS 304
Sterilization Cycle
Solid cycle, Liquid cycle, Liquid with warming cycle, Agar cycle, Solid with drying (for D model), Customized cycle