Empowering Scientific Research Through Analytical Excellence

Pulse Vacuum Tabletop Autoclave AB-400 Series

Pulse Vacuum Tabletop Autoclave AB-400 Series

Based on the new class B of international EN-13060, with 3 pulsating vacuum. Vacuum up to-0.9 Bar (-0.09Mpa). Suitable to sterilize wrap, unwrap, solid, porous, hollow device.

Description
Specification
Technical

Based on the new class B of international EN-13060, with 3 pulsating vacuum. Vacuum up to-0.9 Bar (-0.09Mpa). Suitable to sterilize wrap, unwrap, solid, porous, hollow device.

  • Standard Class B with three times vacuum and drying, the remaining temperature of sterilized instrument is less than 0.2%.
  • LCD display working status, touch keys.
  • The vacuum data can reach -0.8bar, it fits various instrument, including package, unpackaged, solid, hollow, multi-apertures and inset pipes. It makes sure the instrument can be sterilized completely.
  • Fully computer controlled LCD display bright show. The interface is modularized panel for easy operation.
  • It is installed with BOWIE&DICK which measures the penetration of water steam.
  • It has vacuum text which can make sure the ability of texting vacuuming.
  • Fast and independent steam generator.
  • Built-in mini printer which can record the sterilizer information.
  • Safety valve Safety lock door system Pressure or temperature overload protection, Alarm system.
  • Can inactivate HIV, HBV, mad cow virus and bacillus.

 

  • Advanced 16-bit microprocessor, easy to operate. Most suitable for the Dept of Stomatology, Ophthalmology, Surgery and Lab.
  • With BD test, Helix and vacuum test procedures.
  • The jet type of steam generator, ensure the efficient sterilizing..
  • Alarm system for waste water tank, avoid the waste water to enter the sterilizing circulation,.
  • Precise digital display and advanced self-test system ensure you to get the working data freely.
  • Double door-locking system, guarantees the operation and security.
  • Optional inbuilt mini-printer or USB connector to record the process of sterilizing.
  • The new design of water tank on top for easy maintenance.

Model

AB-412

AB-418

AB-424

AB-435

AB-450

Sterilising chamber volume

12L

18L

24L

35L

50L

Chamber Dimensions

ϕ200×370 mm

ϕ250×360 mm

ϕ250×470 mm

ϕ300×500 mm

ϕ340×550 mm

Working pressure

0.22 Mpa

Working temperature

134 ℃

Temperature Range

105-134 ℃

Timer

0-99 min

Heat average

≤ ±1 ℃

Power

1.8KW

2KW

2KW

2.7KW

2.7KW

Dimension(mm)

560×470×400

560×470×400

675×470×400

730×550×510

770×600×540

Net Weight (kg)

43

47

53

93

106

Shipping Package Dimensions (mm)

655×560×495

655×560×495

765×560×495

810×680×650

880×730×700

Gross Weight (kg)

48

52

58

115

130

Electrical Requirements

 

 

 

 

 

 Catalogue No.

 

 

 

 

 

 

Model

AB-412

AB-418

AB-424

AB-435

AB-450

Sterilising chamber volume

12L

18L

24L

35L

50L

Chamber Dimensions

ϕ200×370 mm

ϕ250×360 mm

ϕ250×470 mm

ϕ300×500 mm

ϕ340×550 mm

Working pressure

0.22 Mpa

Working temperature

134 ℃

Temperature Range

105-134 ℃

Timer

0-99 min

Heat average

≤ ±1 ℃

Power

1.8KW

2KW

2KW

2.7KW

2.7KW

Dimension(mm)

560×470×400

560×470×400

675×470×400

730×550×510

770×600×540

Net Weight (kg)

43

47

53

93

106

Shipping Package Dimensions (mm)

655×560×495

655×560×495

765×560×495

810×680×650

880×730×700

Gross Weight (kg)

48

52

58

115

130

Electrical Requirements

 

 

 

 

 

 Catalogue No.

 

 

 

 

 

 